Electrical and electronics measurements miscellaneous
- A 0 – 300 V voltmeter has an error of ± 2% of full scale deflection. If true voltage is 30 V, the range of readings will be
-
View Hint View Answer Discuss in Forum
NA
Correct Option: A
NA
- An 0 – 10 A ammeter has guaranteed accuracy of 1 percent of full scale deflection. The limiting error while reading 2.5 is
-
View Hint View Answer Discuss in Forum
NA
Correct Option: C
NA
- The power in a 3-phase, 3-wire load is measured using two 100W full-scale wattmeters W1 and W2. W1 is of a accuracy class ± 1% and reads 100 W. W2 is of accuracy class ± 0.5% and reads – 50 W. Then the uncertainty in the computation to total power is
-
View Hint View Answer Discuss in Forum
W1 = 100W + 1W;
W2 = – 50 + 0.5W∴ W1 + W2 = 50 ± 1.5 × 100 = ± 3% 50 Correct Option: D
W1 = 100W + 1W;
W2 = – 50 + 0.5W∴ W1 + W2 = 50 ± 1.5 × 100 = ± 3% 50
- The high torque to weight ratio in an analog indicating instrument indicates
-
View Hint View Answer Discuss in Forum
NA
Correct Option: B
NA
- A PMMC instrument produces a deflection of 90° when a current 2 mA passes through its moving coil. The instrument is spring controlled. Supposing the instrument is redesigned with its spring constant as before, and its permanent magnet is replaced with a new permanent magnet which reduces twice the flux density as before, what would be the deflection if a current of 1 mA is passed through the coil? The other features of the PMMC instrument remain the same.
-
View Hint View Answer Discuss in Forum
NA
Correct Option: B
NA