Electrical and electronics measurements miscellaneous
- A piezo electric transducer has the following parameter values:
Crystal capacitance = 10–9 F
Cable capacitance = 2 × 10–10 F
Charge sensitivity = 4 × 10–6 coulomb/cm
If the oscilloscope used for read-out has an input resistance of 1 MW in parallel with C = 4 × 10–10 F, then the voltage sensitivity constant will be
-
View Hint View Answer Discuss in Forum
Total capacitance = Crystal capacitance + Cable capacitance + Oscilloscope capacitance
= (10 + 2 + 4) × 10–10 F
Change sensitivity = 4 × 106 C/cmVoltage sensitivity = Change sensitivity Total capacitance = 4 × 106 = 2500 V/cm 16 × 10–10
Correct Option: A
Total capacitance = Crystal capacitance + Cable capacitance + Oscilloscope capacitance
= (10 + 2 + 4) × 10–10 F
Change sensitivity = 4 × 106 C/cmVoltage sensitivity = Change sensitivity Total capacitance = 4 × 106 = 2500 V/cm 16 × 10–10
- The measurement of a quantity
-
View Hint View Answer Discuss in Forum
NA
Correct Option: C
NA
- The resistance of two coils of a wattmeter are 0-0.1 ohm and 1000 ohms respectively and both are non-inductive. The load current is 20 A and the voltage across the load is 30 V. In one of the two ways of connecting the voltage coil, the error in the reading would be
-
View Hint View Answer Discuss in Forum
rcc = 0.01 ohm, rvc = 1000 ohm.
There are two ways in which the current coil and voltage coil can be connected. The wattmeter reads high by an amount equal to power loss in current coil in one case and in voltage coil in the other. The two possible power losses are
(i) IC² rcc = 20² × 0.01 = 4W
(ii)V² = 30² = 0.9 W rcc 1000
The two percentage errors are
(i)4 × 100 = 2 600 3
(ii)0.9 × 100 = 0.15 600 Correct Option: C
rcc = 0.01 ohm, rvc = 1000 ohm.
There are two ways in which the current coil and voltage coil can be connected. The wattmeter reads high by an amount equal to power loss in current coil in one case and in voltage coil in the other. The two possible power losses are
(i) IC² rcc = 20² × 0.01 = 4W
(ii)V² = 30² = 0.9 W rcc 1000
The two percentage errors are
(i)4 × 100 = 2 600 3
(ii)0.9 × 100 = 0.15 600
- When damping of an instrument is adjusted to enable the pointer to rise quickly to its deflected position without overshooting, in that case the instrument is said to be
-
View Hint View Answer Discuss in Forum
NA
Correct Option: A
NA
- A 50 Hz voltage is measured with a moving iron voltmeter and a rectifier type ac voltmeter connected in parallel. If the meter readings are V1 and V2 respectively and the meters are free from calibration errors, then the form factor of the ac voltage may be estimated as
-
View Hint View Answer Discuss in Forum
Form factor of the wave = rms Value Mean Value
Moving iron Instrument will show rms value. Rectifier voltmeter is calibrate to read rms value of sinusoidal voltage, i.e. with form factor of 1.11.∴ Mean value of the applied voltage = V2 1.11 ∴ Form factor = V1 = 1011V1 V2/1.11 V2 Correct Option: B
Form factor of the wave = rms Value Mean Value
Moving iron Instrument will show rms value. Rectifier voltmeter is calibrate to read rms value of sinusoidal voltage, i.e. with form factor of 1.11.∴ Mean value of the applied voltage = V2 1.11 ∴ Form factor = V1 = 1011V1 V2/1.11 V2